Custom Electronics Systems, Inc.
455 FORUM PARKWAY RURAL HALL, NORTH CAROLINA
27045
PHONE: (336) 969 2411 FAX: 336 969 2412
EMAIL: INFO@DRAFTRANGE.COM
CES 436 Loop Calibrator
FEATURES
Low Cost
Rapid Calibration
Dual Function
Loop Status Indicator
GENERAL DESCRIPTION
The CES Loop Calibrator incorporates a 4 to 20 mA current source and a loop powered simulator in
a convenient to use hand held instrument.
Three internal 9 volt transistor radio batteries provide sufficient power for the CES 436 to source 4 to
20 mA into a 0 to 1000 ohm load. In the two wire mode, operating power is derived from the loop under
test. The CES 436 will operate with a 6 to 35 volt drop across the transmitter.
Failure of theloop status led to illuminate indicates insufficient loop current
due to an open loop, high loop inpedance, loop supply failure(2 wire), or low battery power.
Switch selectable 4, 12, and 20mA settings ensure rapid calibration with
laboratory accuracy in the field. When the 20 mA output is selected , the variable control maybe used to
generate
any output for 4 to 20 mA. |